Heat output
The heat output data is listed in the table below.
Heat outputComponent
Maximum 1500 WBioreactor system, including the rocker
unit, CBCU and pump.
Gas supply
The table below lists the gas supply required for each installed CBCU.
Flow (L/min)
Fast fill mode
Flow (L/min)
Normal mode
Pressure (bar)Gas
3.51.31.0 to 1.5Compressed air/N
2
0.50.21.0 to 1.5CO
2
1.70.71.0 to 1.5O
2
80 ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
4 Installation
4.1 Site preparation
4.1.2 Site requirements

Table of Contents