RequirementParameter
Pollution degree 2Pollution degree of the intended environ-
ment
For satisfactory temperature control, the ambient temperature must be at
least 5°C lower than the desired culture growth temperature.
Note:
The load cells in the rocker are sensitive to changes in temperature. The ambient
temperature should be maintained as constant as possible for reliable weight
measurements.
Note:
Electrical power requirements
The table below specifies the power requirements:
SpecificationParameter
100 to 120 V AC ±10%Supply voltage
200 to 240 V AC ±10%
50/60 HzFrequency
Overvoltage category IITransient overvoltages
1500 VAMax power
EU or US plug, outlet fused or protected
by equivalent circuit breaker.
UK mains cable available on request.
Mains connection
2 m. Extension cables may be used if re-
quired.
Length of mains cable
WARNING
Protective ground. The product must always be connected to a
grounded power outlet.
Noise level
The noise level of the operating bioreactor system is below 65 dBA.
Audio warning signals have a higher noise level than the operating system.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD 79
4 Installation
4.1 Site preparation
4.1.2 Site requirements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ReadyToProcess WAVE 25 and is the answer not in the manual?

GE ReadyToProcess WAVE 25 Specifications

General IconGeneral
BrandGE
ModelReadyToProcess WAVE 25
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals