RequirementParameter
Pollution degree 2Pollution degree of the intended environ-
ment
For satisfactory temperature control, the ambient temperature must be at
least 5°C lower than the desired culture growth temperature.
Note:
The load cells in the rocker are sensitive to changes in temperature. The ambient
temperature should be maintained as constant as possible for reliable weight
measurements.
Note:
Electrical power requirements
The table below specifies the power requirements:
SpecificationParameter
100 to 120 V AC ±10%Supply voltage
200 to 240 V AC ±10%
50/60 HzFrequency
Overvoltage category IITransient overvoltages
1500 VAMax power
EU or US plug, outlet fused or protected
by equivalent circuit breaker.
UK mains cable available on request.
Mains connection
2 m. Extension cables may be used if re-
quired.
Length of mains cable
WARNING
Protective ground. The product must always be connected to a
grounded power outlet.
Noise level
The noise level of the operating bioreactor system is below 65 dBA.
Audio warning signals have a higher noise level than the operating system.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD 79
4 Installation
4.1 Site preparation
4.1.2 Site requirements