Chapter 8 - Replacement Procedures 8-55
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
5.) Reconnect omega bracket: attach 5 screws connecting the omega bracket
6.) Install the MPB module.
7.) Install MPB Front Metal Door.
8.) Refit the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, Right Side
eTower, Upper eTower Front Cover.
9.) Install all accessories.
10.)Turn ON power to the system.
Figure 8-45 BE Omega Bracket Repositioning
• MPB Module Installation Procedure
• MPB Front Metal Door Installation Procedure
• Upper eTower Front Cover Installation Procedure
• Left Side eTower Cover Installation Procedure
• Right Side eTower Cover Installation Procedure
• Lower Front eTower Cover Installation Procedure
• RS Probe Cover Installation Procedure
• Accessories - Replacement Procedures
Perform the checks listed in Back End (BE) Module Replacement Procedure on page 8-217