EasyManua.ls Logo

Gerhardt Vapodest 50s carousel - Titration; Functions Titration Solution; Parameters

Gerhardt Vapodest 50s carousel
83 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
Vapodest 50s carousel
7 System settings
7.3. Titration
In the menu bar select System > Settings > Titration, to open the following
window.
7.3.1. Functions titration solution
Selection of the titration acid and its concentration [mol/l].
- Default: HCL, H2SO4 and other
7.3.2. Parameters
- Titration speed
The titration speed can be inuenced by this factor.
It considers the various pH-changes for various concentrations of the titration
solutions and boric acids. If the factor is set to 1, a titration is done tted to a 2%
boric acid, which is titrated with a respective (0.1 mol/l) titration solution. If another
titration solution is selected, e.g. with a concentration of 0.5 mol/l. there is the
danger of uncontrolled over dosing with same titration functions. Here, the titration
speed can be adapted with the help of the factor.
Rule is: Factor > 1, titration speed is reduced
Factor < 1, Titration speed is increased
Calculation of titration speed
pHx = start continuous pumping in distilling mode
c = Concentration in mol/l
S = Titration speed
- Mode
You can chose between 3 different tables or by selecting "Curve" work without
tables. Detailed information on how to work with titration tables is available in chap.
'7.8. Titration table'.
pHx - 6,1668
c x 6,66
S =