10.6. Manual change of sample - Single determination
The sample glasses of a series can also be fed manually. This option is useful
when glasses are used whose size are not accepted by the carousel.
10.6.1.Modicationsforsingledetermination
Prior to starting with the single determination you have to modify the instrument
in the following way:
1. Take out the guidance cage (see chapter 10.2.2 Change guidance cage).
2. The manual clamping lever is enclosed in the delivery of the Vapodest. It is
located in the tank drawer and is attached to the protection hood for the Z-axis.
To exchange it, pull the quick clamping device towards you (chapter 4.3, pos.
2) and then, insert the manual clamping lever to the stop in the opening.
The digestion tube has to sit tightly after inserting it. Should this not be the
case, then you must push manual tension device deeper .
10.6.2. Start single determination
1. Turn instrument off and on!
2. Press down quick clamping device and put in digestion tube. Make sure to that
you insert the inlet tubing ‘steam’ into the digestion tube at the same time.
3. Check tight tting of the glass and close protective door.
4. Set level sensor control of the tanks to "inactive" (see chapter 7.2.2. Using
external set of tanks), otherwise an error message will be displayed.
5. Open sample series list and start analysis.
6. After terminating each individual analysis press tension lever downwards, take
out sample glass and insert new sample glass.
After the last sample has been determined the message "Analysis over" is shown
on the display. You can now either start a new series or terminate the analysis.