7.8. Titration table
In the Vapodest system a pH-end point determination for the detection of the
measuring value. The functions in the titration table have an inuence on the titra-
tion speed. The lower the N-content of the sample the more careful the additional
amount of titration solution has to be set. The concentration of the boric acid used
is an additional inuence factor. Depending on the method, you have to test seve-
ral functions, to obtain optimal results. Please check the three sample tables for
various boric acid concentrations which are located at the end of this chapter.
7.8.1. Set titration table
From the menu bar select System > Titrator > Titration table, to open the fol-
lowing window- Titration speed.
- Number
You can dene table 1 - 3
- Name
Enter a name for the table
- Delta pH
Indication of the pH-difference - given
- Time for blending (Delay)
During this time a stable pH-measurement has to be done till the next dosing
step (choose between 1 - 99 s).
- Step
Additional amount of titration solution can be chosen between 0,05 - 9,95 ml
Press key "Store", to accept functions. In chapter "8. Create a method" you are
informed how the titration table can be allocated to the various methods.