What to do if the Haemonetics ACP 215 displays DPM OUT OF TOLERANCE?
- SScott TranAug 26, 2025
If the Haemonetics Laboratory Equipment displays DPM OUT OF TOLERANCE: 1. Refer to the system schematic and check, repair or replace all electrical connections associated with the DPM. 2. Measure DPM signal offset (at ambient pressure) at J105 pin 7 on the CPU PCB or check in Diagnostics. If it is not 2.32–2.68 VDC or 1902–2194, replace DPM. Recalibrate DPM in A/D in Utilities. 3. Replace CPU PCB. Using Utilities, calibrate all components under A/D, and centrifuge, and configure the language. Reset protocol parameters (MODIFY PROGRAM) if parameters were different from default values.