4.7 Hilow Motor
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 26 Century+™ Bed Service Manual (man241)
4. Connect the hilow motor power connector (A).
5. If applicable, connect the battery backup (see “Battery Backup Box” on
page 4-8).
6. Plug the bed into an appropriate power source.
WARNING:
Powered bed mechanisms can cause serious injury. Operate the bed
only with persons clear of mechanisms.
7. Test the hilow motor functions to ensure that the bed is working properly.
Adjustment
None