Initial Actions
Chapter 2: Troubleshooting Procedures
Century+™ Bed Service Manual (man241) Page 2 - 5
2
Initial Actions
Use Initial Actions to gather information from operators concerning problems
with the Century+™ Bed. Note symptoms or other information concerning the
problem that the operator describes. This information helps identify the
probable cause.
1. Someone who can explain the problem is available.
Yes No
↓→Go to “Function Checks” on page 2-5.
2. Ask that person to demonstrate or explain the problem. The problem can be
duplicated.
Yes No
↓→Go to “Function Checks” on page 2-5.
3. The problem is a result of improper operator action.
Yes No
↓→Go to “Function Checks” on page 2-5.
4. Perform the “Function Checks” on page 2-5 to ensure proper operation of
the Century+™ Bed.
Function Checks
1. Initial Actions have been performed.
Yes No
↓→Go to “Initial Actions” on page 2-5.
2. The bed is plugged into an appropriate power source.
Yes No
↓→Plug the bed into an appropriate power source. If this solves the
problem, go to “Final Actions” on page 2-7. Otherwise, go to
step 3.
3. The lockouts on the functions are off.
Yes No
↓→Turn off the lockouts and proceed to step 4.
4. The bed functions are not at the limit of their movement.