4.8 Hilow Drive Screw Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 28 Century+™ Bed Service Manual (man241)
Figure 4-8. Hilow Drive Screw Assembly
4. Push the hilow tie rods (B) out of the holes at the head end of the bed.
5. Using the 7/16" socket and ratchet, and the 1/2" open-end wrench, remove
the two bolts (E), nuts (C), and washers (D).
6. Slide the hilow drive screw assembly (F) towards the foot end of the bed,
so that the end of the drive screw (G) comes out of the mounting bracket
hole (H).
NOTE:
You may have to angle the hilow drive screw assembly in order to get the
small roll pin at the end of the drive screw to go through the mounting bracket
hole.
7. Remove the hilow drive screw assembly (F) from the bed.
Replacement
1. Insert the hilow drive screw assembly (F) into the bed so that the end of the
drive screw (G) goes through the mounting bracket hole (H).
m241_123