4.8 Hilow Drive Screw Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Century+™ Bed Service Manual (man241) Page 4 - 29
4
2. Using the 7/16" socket and ratchet, and the 1/2" open-end wrench, secure
the hilow drive screw assembly (F) with the bolts (E), nuts (C), and
washers (D).
3. Insert the hilow tie rods (B) into the holes at the head end of the bed.
4. Secure the hilow tie rods (B) with the cotter pins (A).
5. If applicable, attach the hilow motor to the bed (see “Hilow Motor” on
page 4-24).
6. If applicable, connect the battery backup (see “Battery Backup Box” on
page 4-8).
7. Plug the bed into an appropriate power source.
WARNING:
Powered bed mechanisms can cause serious injury. Operate the bed
only with persons clear of mechanisms.
8. Test the hilow functions to ensure that the bed is working properly.