2.7 The Bed Down Function Does Not Work
Chapter 2: Troubleshooting Procedures
Century+™ Bed Service Manual (man241) Page 2 - 53
2
Figure 2-12. P200 Location
c. Activate the bed down switch.
d. The voltage is 12V DC (+ 3V).
Yes No
↓→Go to step 11.
7. Perform the following:
a. Disconnect the power cable to the hilow motor (B) (see figure 2-13 on
page 2-54).
m241_111