ACM 150
114
10.2.4.5.
Kinked Tube test
Diagnostics> Hardware Service> Kinked Tube Test
Allows the detection of kinked sample tubes by analyzing the Sample Flow pressure. If a
tube is kinked or bent severely and the pressure does not reach the configurable setpoint
Kinked Tube Threshold, a service event indicates the port and the pressure value. The
Kinked Tube Test may be triggered automatically or manually. The time of day and interval
may be chosen for automatic triggering. To trigger the Kinked Tube Test manually, the
user must log in to the console and press the “START.”
Figure 78. Kinked Tube Test
10.2.4.6.
Line Leak Test
Diagnostics>Hardware Service>Line leak test
Any authorized user can perform this test, provided your ACM 150 is configured for
automatic line leak tests. Typically, this test is initiated because one or more lines failed
their automatic test. This test verifies successful corrective actions if services were
performed to resolve the failure. In this mode, you can observe the pressure test of each
line as it occurs.