ACM 150
43
4.
Installation environment
4.1.
Space requirement
No access is required from the rear or either side. The ACM 150 monitor may be against
a wall or set back-to-back with another monitor. Allow a space of 3 inches (8 cm) behind
the rear for air circulation. Front access is required. Leave a clear space of 36 inches (91
cm) in front of the cabinet door.
The approximate overall dimensions of the cabinet are:
Height: 60 inches (152 cm)
Width: 40 inches (102 cm)
Depth: 26 inches (66 cm)
Weight: 800 pounds (363 kg)
4.1.1. Temperature and protection
Install indoors in a heated and air-conditioned environment. The recommended
temperature range is 41°F to 86 °F (5°C to 30°C). The standard cabinet is designed to
meet an IP 20 rating.
4.1.2. Electrical area
The ACM 150 monitor is designed to be in a non-hazardous or general-purpose electrical
area.
Refer to Technical Specifications on Technical Specifications.
4.2.
Utility requirements
The ACM 150 monitor requires the following utilities:
Line power: 115 V 50/60 Hz 13 A or 230 V 50/60 Hz 6 A (setup at factory)
Normal idle power consumption: ~8 A @ 115 Vac, ~4 A @ 230 Vac
Average power consumption: ~14 A @ 115 Vac, ~7 A @ 230 Vac
Peak power consumption: ~20 A @ 115 Vac, ~10 A @ 230 Vac
Clean, Dry Air (CDA): 90 psig, 12 SCFM, 3/8-inch OD tube
Nitrogen: 5 psig, ~8 liters/minute, 1/4-inch OD tube
Exhaust: 15 cfm, DN40 (1-1/2 inches)
Main supply voltage: ±10% of nominal voltage
Overvoltage: Category II