ACM 150
149
Appendix E.
Technical Specifications
Performance Specifications
Analysis Method
Continuous scan FT-IR analyzer
(FT-IR = Fourier transform infrared)
Gas Cell Path
5.0 m path length
Gases Monitored
Organic, PFC, CFC, HFC, Metal-Organic,
NF3, and a wide variety of other inorganic
gases
Lower Detection Limit
0.1 to 2.0 ppm for 5.0 m path cell, gas-
dependent
Scanning Rate
15 seconds per point
Sample Point Capacity
Available in 10, 20, 30, and 40 points
configurations composite sampling up to 4
points
Data Archiving
One day for all spectra, 365 days for alarms
Applicable Standards
EN 50270, EN 61010, UL 61010
Installation
Indoor only
Altitude
Up to 2,000 m (6,562 ft.)
Temperature range
5°C to 30°C (41°F to 86°F)
Relative humidity
80% at temperatures up to 31°C (88°F),
decreasing linearly to 50% RH at 40°C
(104°F)
Main supply voltage
±10% of the nominal voltage
Overvoltage
Category II
Pollution degree
2
Ingress protection
IP 20
Facilities Requirements
Sample Line Tubing
1/4” ID x 3/8” OD polypropylene
Sample Line Inlets
3/8” connector
Electrical Supply
230 Vac, 10 A or 115 Vac, 20 A
Purge Gas, N2
5 - 10 psi, 10 L/min
Venturi Requirements
90 - 120 psi, 320 L/min
Exhaust Pipe
1.5” NPT
Overall Dimensions
66 X 34 X 25 inches (H X W X D)
1,676 X 864 X 635 mm (H X W X D)
Weight
Operating
349 kg / 770 lb
Shipping
470 kg / 1036 lb