134
11.2.
Resolving problems
Table 8 Lists service messages that the ACM 150 monitor software might generate.
Troubleshoot the problems by following the recommended actions. Use ‘refer to’ for
guidance, additional information, and step-by-step procedures.
Table 8: Service Request Messages
Possible line leak
problems in line xx
Test the line and its check valve
for leaks
Line leak tests
(optional)
Possible line leak
problems in lines (all)
Check the venturi regulator &
air supply. Check the line leak
•
Line leak test setup
procedure
Energy (ZPD) value too
low (or too high)
Evaluate the optical path
Evaluating the FTIR
optical path
No data from the bench
malfunction. Evaluate the
Evaluating the FTIR
optical path
No active scans are
defined.
Set up an Order of Scan
communication with
Verify that the cables and power
cords are connected to the
Interconnecting cables
Cannot update relays
Verify that the cables and power
cords are connected to the relay
Interconnecting cables
Cannot load factory
background spectra
Check for missing files or path
errors (should not occur after
Contact Honeywell
addresses out-of-range
Too little reaching detector
Evaluating the FTIR
optical path
Incorrect data length
(short/long scan)
Poor connection at FTIR data
cable
Evaluating the FTIR
optical path