Honeywell ACM 150 - Figure 36. Normal Scan

Honeywell ACM 150
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACM 150
65
Figure 36. Normal Scan
Point
NOTE
Area Number, Port Number, and Point
Number all correspond to the inlet port
number, 1 - 40, on top of the ACM 150
monitor. This is where the sample tube
connects the ACM 150 monitor to the
remotely sampled locations.
Those sampled locations are always
identified with the same 1 - 40 number
but may be further identified using an
area name.
This displays the name the user
designates for the sample area,
corresponding to the area number.
Since the area number is unlikely to have
any meaning to the operator, the ACM
150 monitor software allows you to enter
the name or equipment ID #.
Status
Status for each active gas on the display
OK: Indicates the gas concentration is
lower than the warning level set point.
Warn: Indicates the gas concentration
is above LDL
Alarm 1: Indicates the gas
concentration exceeds the Alarm 1
level but is lower than the Alarm 2
level.
Alarm 2: Indicates the gas
concentration exceeds the Alarm 2
level set point.