140
Fault Category Event Level
Event
Code
Displayed Message Remark
Scan Fault
System Down 41 Scan Fault - No background spectra available
Require Service 42 Scan Fault - No factory background spectra available
Require Service 6
Scan Fault - Cell pressure failed to reach # psi.
(Cell pressure on Point# during Fill: # psi)
# Point area-id
Require Service 7
Scan Fault - Sample pressure failed to reach # psi.
(Sample pressure on Point# during Fill: # psi)
# Point area-id
Overrange Fault
System Down 44 Overrange spectra detected over #. # Alarm threshold
Require Service 44 Overrange spectra detected over #. # Warn threshold
Diagnostics
Information
Require Service 4
Diag Info - Possible kinked tube detected.
(Sample pressure on point#: # psi)
# Sample valve number
# Pressure
Information 103 Diag Info - Kinked Tube Test is started
Information 104 Diag Info - Kinked Tube Test is completed
Require Service 3
Diag Info - Possible line leak detected.
(Sample pressure on Point #,#, …,#: # psi)
Diag Info - Possible line leak detected.
(Sample pressure on Point #: # psi)
# … # Sample valve number
# Pressure
Information 105 Diag Info - Line Leak Test is started
Information 106 Diag Info - Line Leak Test is completed
Require Service 43 Diag Info - Need to acquire new background spectra
Information 100 Diag Info - " " is logged in. "Username"
Information 101 Diag Info - " " is logged out. "Username"
Information 102 Diag Info - Operation mode is changed to " " mode. "Operation mode"
Information 107 Diag info - Succeed to create MTD
Require Service 108 Diag info - Failed to create MTD
Information 109 Diag info - Succeed to create TWA
Require Service 110 Diag info - Failed to create TWA
System Information
Information 111 Diag info - Succeed to import configuration (*.zip)
Require Service 112 Diag info - Failed to import configuration (*.zip)