CalibrationGasConcentration: 25%LELto105%LEL
AlarmThresholdLimits:  15%LELto100%LEL  
  (Alarm 1 and Alarm 2)
Poisoning:   Thecartridgemaybecomeinactiveafterextensive 
  exposure to silicones, halogenated hydrocarbons, heavy  
  metals or sulfur compounds.
OxygenDeciency:  DeletionofOxygenasaconsequenceofdisplacementby
 ammablegascanresultinthegasreadingreturning 
 tozero.
StorageConditions:  0to25°C
   30%to70%rH
   80kPato110kPa
Thissub-sectionprovidesthespecicationsforeachoftheaccessoriesavailableforApex.
A.3.1  Certied Junction Box
TheCertiedJunctionBoxprovidesaremotemountingpointforthesensor.Itisdesignedtooperate
withthesensorcartridgesspeciedinthetablesinthissection.
 
Environmental
Operatingtemperature:   -40to65°C(-40to149°F)
Operatinghumidity:   0-99%rh
Storage (excluding cartridge)
Temperature:   -40to65°C(-40to149°F)
Humidity:   0-99%rh
EMC
MeetstherequirementsofEN50270(type2apparatus),EMC-ElectricalApparatusfortheDetection
and Measurement of Combustible Gases, Toxic Gases or Oxygen.
Enclosure
Physical:   TheCertiedJunctionBoxhas:
       One threaded hole in the base for mounting the 
  CertiedSensor-M20(European)or∫”NPT(North 
   American versions)
   Twothreadedholesinthebasetoaccepttwocertied 
  M25cableglands(European)ortwo∫”NPTconduit 
  ttings(NorthAmericanversions).
       Provision is made to attach a tag carrying an ID number. 
   TheenclosurematerialfortheCertiedJunctionBoxis 
     electro-polished grade 316 stainless steel (to BS3146  
     ANC 4B FM/BS3146 ANC 4C FM). 
APPENDIX A - SPECIFICATIONS