4. OPERATION
4.6 FAULT DIAGNOSIS
Apex system faults can be those displayed on the Transmitter Unit screen or be of a general type.
Thissectiondescribes:
Error messages displayed on the Transmitter Unit LCD screen.
These are either fault or warning types. They can normally be corrected at the Transmitter
UnitorCertiedSensor.
Errors that are not diagnosed by the system itself but need
some action.
ForexampletheCertiedSensorreadslowwhengasisapplied.Torectifythisfaultthe
control system span needs to be adjusted.
Signal ranges and fault conditions on the 4-20mA output.
How to clear latched alarms
4.6.1 Displayed Error Messages
TheerrormessagesontheLCDscreenaresimilartothefollowing:
F:No Cartridge
FAULT
Panels flash
alternately
until
fault cleared
The following table lists all the displayed Fault (F) and Warning (W) messages.
Message Cause/Remedy
W: Sensor Failed
F: Sensor Failed ACertiedSensorfaulthasdeveloped.
ChecktheconnectionsbetweentransmitterandCertiedSensor.
Re-calibratetheCertiedSensor.
IffaultremainsreplacethecompleteCertiedSensorassembly.
F: No Cartridge TheCertiedSensorreportsthereisnocartridgetted.
Checkthatthecartridgeisttedcorrectly.
Ifnocartridgepresent,trequiredcartridge.
F: Wrong Cartridge ThewrongtypeofcartridgehasbeenttedintheCertiedSensor.
Replace the cartridge with correct type for the target gas.
F: Cartridge Failed The cartridge has failed.
Replace the cartridge.
W: Temperature Error Thecartridgeisbeingoperatedattemperaturesoutsideitsspecied
F: Temperature Error temperature range.
Identify the cause of temperature problem.
FitsuitableaccessoriesorrelocatetheCertiedSensorasappropriate.
IffaultpersistsreplacetheCertiedSensor.