EasyManua.ls Logo

Honeywell Searchpoint Optima Plus - 7.5 Bump Test (Gas Challenge) with Remote Gassing Cell (RGC)

Honeywell Searchpoint Optima Plus
97 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAN0905_Issue 7_07/17 Searchpoint Optima Plus with optional HART
®
output
Operating Instructions
2108M0550
34
7.5 Bump Test (Gas Challenge) with Remote Gassing Cell (RGC)
Usingatestgasof50%v/vmethaneintheRemoteGassingcell(RGC),agasresponsewillbegenerated
from most Searchpoint Optima Plus units. The reading depends on the target gas calibration of the unit.
The RGC is a tool to check for a gas response and is not precise; therefore variability in response between
units should be expected.
Caution: Gas readings will be generated during this procedure. Take appropriate action at the control
and monitoring system to inhibit signals.
Caution: Maximum ow rate is 0.3 litres per minute to avoid pressurising the gas in the RGC.
Note:ThisprocedureonlyappliesiftheSearchpointOptimaPlushasaRemoteGassingCell(RGC)tted.
TheRGCisafactoryttedoption.
Note:Duringtestingthe4-20mAoutputcanbemonitoredeitheratthecontrolroomorlocallywitha
multimeter. When connecting or disconnecting a multimeter, power off the Searchpoint Optima Plus,
otherwise the unit will indicate a latched warning that the 4-20 mA loop was broken. If this occurs it can be
cleared by re-cycling the power.
Remote Gassing Cell
High Concentration
(v/v cell) Nozzles
1. Inspect the gassing pipes to the RGC. Ensure that the exhaust is clear and vents well away from the
Searchpoint Optima Plus, and does not cause a hazard.
2. Apply50%v/vmethaneatarateof0.3litresperminutefor~30-60s.(Forlonggaspipes>~10m,
allow a longer time for the gas to reach Searchpoint Optima Plus).
3. Check that the Searchpoint Optima Plus responds to the test gas.
Note:Ifthegasreadingisnotacceptablefortheapplication,re-zerotheSearchpointOptimaPlusand
repeatthetest.TheSHC1HandheldInterrogator,theHART
®
communication or the XNX Universal
Transmitter is required to re-zero the Searchpoint Optima Plus.
4. Flush the RGC with zero gas until the gas reading has returned to zero.
7. Maintenance

Table of Contents

Other manuals for Honeywell Searchpoint Optima Plus

Related product manuals