EasyManua.ls Logo

Honeywell Searchpoint Optima Plus - 13.5 Bump Test (Gas Challenge) with Remote Gassing Cell (RGC)

Honeywell Searchpoint Optima Plus
97 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAN0905_Issue 7_07/17 Searchpoint Optima Plus with optional HART
®
output
Operating Instructions
2108M0550
64
9. RemovetheGassingCoverandwaitforthetestgastoclear(~30s).Ensurethatthegasreadinghas
returned to zero. Press OK to release the 4-20 mA output.
10. Return to the Welcome screen.
13.5 Bump Test (Gas Challenge) with Remote Gassing Cell (RGC)
Usingatestgasof50%v/vmethaneintheRemoteGassingcell(RGC),agasresponsewillbegenerated
from most Searchpoint Optima Plus units. The reading depends on the target gas calibration of the unit. The
RGC is a tool to check for a gas response and is not precise; therefore variability in response between units
should be expected.
Caution: Gas readings will be generated during this procedure. The 4-20 mA output will be inhibited
automatically. If the procedure is aborted, ensure that the unit is not left in Inhibit mode.
Caution: Maximum ow rate is 0.3 litres per minute to avoid pressurising the gas in the RGC.
Note:ThisprocedureonlyappliesiftheSearchpointOptimaPlushasaRemoteGassingCell(RGC)tted.
TheRGCisafactoryttedoption.
Remote Gassing Cell
High Concentration
(v/v cell) Nozzles
1. Inspect the gassing pipes to the RGC. Ensure that the exhaust is clear and vents well away from the
Searchpoint Optima Plus, and does not cause a hazard.
2. Select Device Setup.
3. Select Calibrate then Bump Test.
4. Select Start Bump Test and press Enter.
5. Thedisplayshowsaconrmationmessage,pressOK. The 4-20 mA output is now inhibited.
6. Apply50%v/vmethaneatarateof0.3litresperminutefor~30-60s.(Forlonggaspipes>
~10m,allowalongertimeforthegastoreachSearchpointOptimaPlus).
7. Check that the Searchpoint Optima Plus responds to the test gas. The live gas concentration is shown
on the screen. When ready to continue press OK.
8. FlushtheRGCwithzerogasuntilthegasreadinghasreturnedtozero.PressOK to release the
4-20 mA output.
9. Return to the Welcome screen.
13. Maintenance using HART
®
Communications

Table of Contents

Other manuals for Honeywell Searchpoint Optima Plus

Related product manuals