85
18.2.7 CCCF (China)
CerticateNumber: CE031155
ExdIICT4/T5
18.2.8 IECEx
CerticateNumber: BAS13.0069X
18.3 Performance Approvals
18.3.1 ATEX
CerticateNumber: BVS03ATEXG016X
Special conditions for safe use
Thefollowingspecialpropertieshavetobeconsideredatoperationoftheremotesensor:
• SeeEC-typeexaminationcerticateBaseefa13ATEX0296X
• The parameter “warning” shall not be set to values between 3 mA and 5 mA
• Searchpoint Optima Plus shall be calibrated using the target gas or vapour
• TheModbus-interfaceisnotsubjectofthissupplementtotheEC-typeexaminationcerticate
• TheHART-interfaceisincludedinthissupplementtotheEC-typeexaminationcerticatewith
respecttotheuseforconguration,commissioning,testandmaintenanceoftheremotesensor
• DonotusethelightringindicationofthejunctionboxHALOforsafetyrelatedpurposes
18.3.2 FM
Certicatenumber:3015165
This Approval does not include or imply Approval of apparatus to which the subject instrumentation may
be connected. In order to maintain an FM Approved system, the apparatus to which this instrument is
connected must also be Approved by FM Approvals.
AspartofthisApproval,itwasveriedthatoptionalcommunicationfunctionsofthisgasdetection
instrument while operating at the maximum transaction rate do not adversely affect the gas detection
operation and functions of the instrument. This Approval, however, does not include or imply Approval
of the communications protocol or functions provided by the software of this instrument or of the
communications apparatus or software connected to this instrument.
18. Certication and Approvals