86
18. Certication and Approvals
18.3.3 Russian Pattern Approval (Metrology) - XTC Version
NOTE: For instruments manufactured in the Russian Federation, calibration is available only for these
gases: methane, ethane, propane, butane, ethylene.
Thisapprovalcoversthefollowinggases:
as Name Measuring Range (%LEL) Measuring Range (expressed in %v/v)
Methane 0 to 100 0 to 4.4
Ethane 0 to 100 0 to 2.5
Propane 0 to 100 0 to 1.7
Butane 0 to 100 0 to 1.4
Heptane 0 to 100 0 to 1.1
Hexane 0 to 100 0 to 1.0
Methanol 0 to 50 0 to 3.75
Toluene 0 to 100 0 to 1.1
o-Xylene 0 to 20 0 to 0.2
p-Xylene 0 to 18 0 to 0.2
Pentanes 0 to 100 0 to 1.4
Octane 0 to 50 0 to 0.4
Butadiene 0 to 100 0 to 1.4
Isobutane 0 to 100 0 to 1.3
Methane - 0 to 100
Ethylene 0 to 100 0 to 2.3
Propylene 0 to 100 0 to 2.00
Benzene 0 to 100 0 to 1.2
Sterol 0 to 27 0 to 0.3
PleaserefertotheMetrologycerticateforSearchpointOptimaXTCforfulldetails.
18.4 Marine Approvals
Searchpoint Optima Plus complies with the EC Marine Equipment Directive (MED).
0062/XX
Note:XXrepresentstheyearofmanufacture
TypeApprovals: BureauVeritas
American Bureau of Shipping
Det Norsk Veritas
Lloyd’sRegister