EasyManuals Logo

Honeywell Searchpoint Optima Plus User Manual

Honeywell Searchpoint Optima Plus
97 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
MAN0905_Issue 7_07/17 Searchpoint Optima Plus with optional HART
®
output
Operating Instructions
2108M0550
53
12.2 First Time Switch On
Note:SearchpointOptimaPlusissuppliedcalibratedandreadyforuse.Itdoesnotrequirecalibration
during commissioning.
Note:Duringcommissioningthe4-20mAoutputcanbemonitoredeitheratthecontrolroomorlocally
with a multimeter. When connecting or disconnecting a multimeter, power off the Searchpoint Optima Plus,
otherwise the unit will indicate a latched warning that the 4-20 mA loop was broken. If this occurs it can be
clearedusingtheHART
®
communications.
1. Beforeapplyingpowertotheunit,ensurethattheeldwiringiscorrectandallelectricalconnections
are in accordance with Section 4.
2. Applypower.CheckthatthesupplyvoltagetotheSearchpointOptimaPlusisbetween18and32V.
3. Monitorthe4-20mAoutputduringthestartupsequence,whichshouldbeasfollows:-
Current output Duration
< 1 mA < 5 s
2 ± 0.2 mA < 60 s
4 ± 0.2 mA ongoing
Note:Ifgasispresent,theunitwillindicateagasreading
Note:Iftheoutputrevertsto3mAor<1mA,aWarningorFaultispresent.Followtheinstructionsbelowto
investigate and clear a warning or fault (see Clear Faults or Warnings).
Searchpoint Optima Plus is factory calibrated and it is not necessary to calibrate or gas test it during
commissioning.Howeveriflocalregulationsorsitepracticesrequirethatagastestisdonetocheckthat
the unit is functioning, please see Chapter 13 Maintenance, Bump Test for instructions.
12.3 User Conguration
12.3.1 Change the Password
ItisrecommendedthatthepasswordforLevel1accessshouldbechangedfromthedefaultsetting.
Thepasswordis8digitalphanumeric(casesensitive).Afull8digitpasswordshouldbeused.
1. Select Device Setup.
2. Select User Login.
3. Select Level 1.
4. Enter the password 00000000 (eight zeroes) and press Enter.
5. Thedisplayshowsaconrmationmessage,pressOK.
6. Select User Conguration.
7. Select Change Password.
8. Thedisplayshowsaconrmationmessage,pressOK.
12. Commissioning using
HART
®
Communications

Table of Contents

Other manuals for Honeywell Searchpoint Optima Plus

Questions and Answers:

Honeywell Searchpoint Optima Plus Specifications

General IconGeneral
BrandHoneywell
ModelSearchpoint Optima Plus
CategoryGas Detectors
LanguageEnglish

Related product manuals