FULLTEST3
EN - 10
3. PREPARATION TO USE
3.1. PRELIMINARY CHECKS
This instrument was checked both mechanically and electrically prior to shipment. All
possible cares and precautions were taken to let you receive the instrument under
perfect conditions. Notwithstanding we suggest you to check it rapidly (possible
damages may have occurred during transport – if so please contact the local
distributor from whom you purchased the item). Make sure all standard accessories
according to Packing List are included. Should you have to return back the
instrument for any reason please follow the instructions mentioned in § 3.4
3.2. POWER SUPPLY
Earthed plugs must energize the instrument. To avoid any risk the instrument does
not allow to effect measurements when there is not such a connection (see § 4.2).
CAUTION
The instrument include EMC/EMI filters which could trip RCD
with 30mA nominal current. It is recommended to power supply
the instrument through a mains protected by a RCD of 100mA (or
higher) nominal current
3.3. STORAGE
In order to guarantee precise measurement, after a long storage time, wait for the
instrument to come back to normal condition (see § 10.6).