EN - 32
6.1.3. Anomalous situations
The following specific information can be shown on the display during measurement
Information displayed Description
CHECK CALIBRATION
Measurement result is negative probably
because of shorter test leads than calibrated
(negative value is higher than 5 digits).
Calibrate test leads again!
EXTERNAL VOLTAGE
x External voltage higher than 3 V is applied
between two R
PE or between two SENSE test
terminals (measurement is not running) or
higher than 10 V (measurement is running).
x External voltage higher than 5 y30 V is applied
between any R
PE or SENSE test terminal and
GND.
Remove external voltage!
LIMIT OUT OF RANGE
Calculated limit value is < 1 (EN60204 SET Z
limit mode).
FUSE F4!
Fuse F4 is blown.
ERROR 1!
Internal fuse may be blown!
The fuse is not customer replaceable, send the
tester into a service department.
MEAS. TIME > 5MIN
CHECK TIMER
The Timer is set to a value higher then 5
minutes with 25A test selected.
The test with 25A current allows to set the timer
up to maximum 5 minutes