FULLTEST3
EN - 39
8. Test result will currently be displayed in green colour (result higher than or equal
to set limit value) or in red colour (result lower than set limit value). Final result will
be equipped with green
symbol and with beep-beep sound (result OK) or with
red
symbol and with longer beep sound (result not OK)
9. Save the test result by pressing the SAVE key (see § 7.1)
6.3.1. Anomalous situations
The following specific information can be shown on the display during measurement
Information displayed Description
EXTERNAL VOLTAGE
x External voltage higher than 10 V AC approx.
is applied between positive and negative test
terminals (measurement is not running) or
higher than 50 V AC approx. (measurement is
running).
x Negative external voltage higher than 10 V DC
approx. is applied between positive and
negative test terminals (measurement is
running).
Remove external voltage!
DISCHARGING!
External capacitor (or internal) that was charged
during the measurement is discharging.
Wait until the message disappears! Do not
disconnect test leads until the message is
present!