FULLTEST3
EN - 7
2. GENERAL DESCRIPTION
The FULLTEST3 is a measurement instrument used for final inspection of electrical
equipment of machines, control cabinets, switchgears as well as other devices
complying with IEC/EN60204-1 and IEC/EN61439-1 standards. The followed
operative instructions are relative to the IEC/EN60204-1 guideline.
2.1. FUNCTION DESCRIPTION
x Continuity of protective conductor
¾ 2-wire or 4-wire measurement method.
¾ Compensation of test leads in case of 2-wire
measurement.
¾ Open-circuit test voltage 6 V AC approx.
¾ Test current 200 mA and 25 A AC.
¾ Limit value adjustable, visual and acoustic warning in
case of exceeded value
EN61557-4
EN61439-1-§10.5.2
EN60204-1-§18.2.2
EN60598-1
EN60335-1-§27.5
EN60335-1-§A.1
EN50106
EN60950
x Insulation resistance
¾ Test voltage 100V, 250V, 500V and 1000V DC
¾ MAN (manual) mode.
¾ TIMER mode.
¾ AUTO mode.
¾ Limit value adjustable, visual and acoustic warning in
case of exceeded value.
EN61557-2
EN61439-1-§11.9
EN60204-1
EN60598-1
x Dielectric test
¾ Adjustable test voltage 250 V up to 5100 V AC.
¾ Trip out current adjustable 1 y 110 mA, visual and
acoustic warning in case of exceeded limit value.
¾ Display and trip out based on real or apparent current.
¾ MANUAL mode
¾ RAMP 75% mode (predefined automatic rising of test
voltage).
¾ RAMP 50% mode (predefined automatic rising of test
voltage)
¾ BURN mode
¾ PULSE mode
¾ Protection against unauthorised use (safety measure).
¾ Red warning lamp connector (safety measure).
¾ Safety input connector (safety measure).
EN61439-1-§9.1
EN60204-1-§18.4
EN60598-1
EN60335-1-§13.3
EN60335-1-§A.2
EN50191
x Loop impedance measurement
¾ ZL/N, ZL/L and ZL/PE measurement.
¾ Voltage range 100 y 460 V.
¾ IPSC calculation.
¾ Limit value adjustable, visual and acoustic warning in
case of exceeded value.
EN60204-1-§18.2
EN61557-3
x Phase sequence
¾ Mains voltages U
L1/2, UL2/3, UL3/1 simultaneously
displayed.
EN61557-7