Checking the measuring accuracy
With the help of the FieldVerification Kit, you can check the error of
measurement of the FluidControl Unit FCU on site based on a fluid defined
as contaminated. Based on these results, you can decide whether you have
the FCU recalibrated or extend the period of its use.
The part no. for the FieldVerification Kit can be found in the accessories list.
The use of the FCUFieldVerification Kit does not eliminate the need for
factory recalibration.
FCUcalibrating
Recalibrate the FCU according to ISO 9000 standard.
We recommend a recalibration of the FCU at least every 3 years.
For calibration, contact one of the following HYDAC national subsidiaries,
see "Finding Customer Service / Service stations" chapter on page 111.