Safety information
Proper/Designated Use
Claims for defects or liability, regardless of the legal foundation, do not apply
with incorrect or improper installation, commissioning, usage, handling,
storage, maintenance, repair, use of unsuitable components or other
circumstances for which HYDAC is not responsible.
HYDAC is not responsible for the installation, integration, selection of
interfaces to / into your system nor for the use or functionality of your system.
Use the product only for the application described below.
The FluidControl Unit FCU is a portable service unit for taking short-term
measurements of the solid particle contamination, temperature und %
saturation level in hydraulic systems.
Proper/designated use of the product extends to the following:
Observing all the instructions contained in these operation and
maintenance instructions.
Ensuring the execution of all inspection and maintenance work.
NOTICE
Impermissible operating conditions
The FCU will be destroyed
Use the FCU with mineral oils or mineral oil-based raffinates whose
flash point is higher than 55°C / 131°F.
Flush the FCU with low-viscosity mineral oils or mineral oil-based
raffinates (e.g. diesel) whose flash point is higher than 55°C / 131°F.
Never flush the FCU with universal thinners or other degreasing fluids.
Observe the permissible viscosity range (up to ISO VG 68):
10 - 350 mm²/s / 46 - 1622 SUS.
Only operate the FCU in periodic intermittent duty (S3) with a relative
duty cycle of 40%
(S3 as per DIN EN 60034 / VDE 0530).
Never use the FCU without a suction strainer.
FCU1310
en(us)
Page 13 / 124
BeWa FCU1310 V300 3371346f en-us 2017-01-27.docx 2017-01-27