Preparing the FCU for measurement
Selecting the measurement location
Select the measurement location so that the sample measured comes
from a turbulent location, with a good flow. For example on a pipe
bend. This ensures that a typical sample is analyzed.
If the FCU is installed near the measurement point, avoid delayed
measurement results and sedimentation (particle deposits in the line).
While installing the inlet hose INLET, make sure that no siphon results.
FCU1310
en(us)
Page 38 / 124
BeWa FCU1310 V300 3371346f en-us 2017-01-27.docx 2017-01-27