Identifying and interpreting status/error messages
Identifying and interpreting status/error messages
The following error codes are possible. You will find these error codes in the
measurement file.
code
Sensor / equipment is working.
Sensor / equipment continues to
work.
WARNING is automatically reset
by the FCU.
Sensor / equipment status is
"fault".
Restart the FCU by switching it off
and then on again.
The sensor or equipment is faulty.
Contact the HYDAC Service
department.
flashing code
code
-
FCUno display
no function
Check the power
supply to the
FCU.
Contact the
HYDAC Service
department.
-
Gree
n
FCU ready for
operation
You can make
further
measurements.
0
Red
The FCU is below
its measurement
range < ISO 9/8/7.
It is currently not
possible to
determine the
cleanliness of the oil
or its flow rate.
You can make
further
measurements.
2
FCU1310
en(us)
Page 99 / 124