INFICON Warning and error messages | 7
Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 65 / 104
No. Message Possible error sources Fault rectification
E25 Remove the sniffer tip
from the opening of the
calibrated leak!
The sniffer tip is in the calibration
opening of the ECO-Check.
Remove the sniffer tip.
The light barrier of the ECO-Check is
soiled.
Blow out the calibration opening with
fresh air or clean it with a cotton
cloth.
W28 Real-time clock was
reset! Please enter
date and time.
CPU-board MC50 was replaced. Enter the date and time, see
Miscellaneous [}32].
The battery on CPU-board MC50 is
defective.
Contact INFICON customer service.
W29 Voltage 24V on the
audio output is too low!
Fuse F6 on the wiring board is blown. Replace the fuse.*
The speaker is defective. Replace the speaker.*
E30 Sensitivity too low The sensor in the Transpector is
defective.
Contact INFICON customer service.
W31 Factor K1 outside of
range (0.9 to 1.1)!
Other interfering gases, not
cyclopentane or isopentane, were
detected, e.g. alcohols.
For new calibration of IGS, see
Suppress interfering gases [}48].
W34 (The flow rate has been
reduced by more than
30% since the last
calibration. The
warning goes out if the
flow change is less
than 20% again.)
The sniffer line is incorrectly
connected.
Check the connection.
The sniffer line filters are clogged. Change the filter of the sniffer line,
see Replacing the filter inserts of the
capillary filter and the water
conservation tip [}80]. Confirm the
work, see Calling up and managing
maintenance information [}71].
Perform new calibration of device,
see Calibrate [}43].
E38 Capillary broken!
(>60s) exceeded. The
pumps are switched off
to protect the
cathodes.)
The upper limit value is set too low. Set the upper limit value for the flow
higher, see Vacuum & access control
[}36].
The capillary is broken or leaking. Replace the sniffer line.
Perform new calibration of device,
see Calibrate [}43].
Replace the multi-function cable of
the sniffer line.Replace the multi-
function cable of the sniffer line.*
Perform new calibration of device,
see Calibrate [}43].