7 | Warning and error messages INFICON
68 / 104 Ecotec E3000-Operating-instructions-kina22en1-28-(2211)
No. Message Possible error sources Fault rectification
W63 TSP parameters
inconsistent! Please
check!
The Transpector was replaced. Check the Transpector parameters.
Contact INFICON customer service.
The wiring board was replaced.
The EEPROM on the wiring board is
defective.
W64 There are warnings
pending!
Confirmed but not yet valid warnings
are repeated every two hours or at
every new start-up.
Remedy the cause of the warning.
Deactivate the warning repetition,
see Calling up and managing
maintenance information [}71].
W65 TSP serial number
inconsistent! Please
check!
The Transpector was replaced. Contact INFICON customer service.
The wiring board was replaced.
The EEPROM on the wiring board is
defective.
W66 ECO-Check brand-
new!
A new ECO-Check calibrated leak
was connected.
Enter the serial number and the code
of the calibrated leak, refer to the
installation manual of the calibrated
leak.
W67 ECO-Check expires on
DD.MM.YYYY!
The validity of the ECO-Check gas
store expires in three months.
Order an ECO-Check gas store.
W68 ECO-Check expired! The validity of the ECO-Check gas
store is expired (2years operation or
older than 3years).
Replace the ECO-Check gas store,
see installation manual of the ECO-
Check.
W70 All EEPROM
parameters of the
ECO-Check lost!
The EEPROM in the ECO-Check
calibrated leak is empty or defective.
Replace the ECO-Check calibrated
leak, see installation manual of the
ECO-Check.
W71 No communication with
ECO-Check!
The ECO-Check calibrated leak
cannot be addressed by the basic
unit.
Check the connection to the ECO-
Check calibrated leak.
No ECO-Check calibrated leak
connected.
If the problem still exists: Contact
INFICON customer service.
W72 No communication with
sniffer line!
The sniffer line cannot be addressed
by the basic unit.
Check the connection of the sniffer
line with the basic unit (disconnect
and reconnect; if possible try a
different sniffer line). If the problem
still exists, contact INFICON
customer service!
E73 Unsuitable sniffer line! The SL3000XL of Protec P3000 was
connected accidentally.
Connect the correct sniffer line.
W74 IC1000 SW version is
obsolete!
A SW update of the IC1000 is
required for full functionality.
Please contact INFICON Service.