INFICON Warning and error messages | 7
Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 67 / 104
No. Message Possible error sources Fault rectification
E48 “Emission failure
(The emission failed
during operation.)
The pre-pressure is too high.
The Transpector is defective.
Confirm the error message and
restart the device. If the problem still
exists: Contact INFICON customer
service.
W49 No emission with first
cathode
Start-up of the Emission failed. The
device has switched to the second
cathode.
You can continue to measure but
need to have the cathodes checked.
E50 Control of turbo pump
or electronic system!
There is an error in the control of the
turbo molecular pump.
Contact INFICON customer service.
E51 No communication with
the turbo controller!
Wiring board defective
CPU board MC50 defective
Contact INFICON customer service.
W57 Too many EEPROM
write cycles!
Continuous change of gas
configuration or I*Guide-Program via
interface.
Use the alternative commands. (e.g.
BEF749 instead of BEF750)
W58 Measuring parameters
inconsistent! Please
check!
The gas library was changed by a
software update.
Perform new selection of the gases
to be measured. (The affected gas
number is displayed.)
W59 Overflow of EEPROM
parameter queue!
EEPROM defective. Contact INFICON customer service.
W60 All EEPROM
parameters lost! Check
your settings!
A new wiring board was installed. All settings are reset to factory
settings. Make your settings again.
If the message occurs constantly
during running up, the EEPROM on
the wiring board is defective.
Contact INFICON customer service.
W61 0 EEPROM parameters
initialized!
A software update introduced new
parameters. The new parameters are
listed in the warning.
Confirm the warning.
If the message occurs constantly
during running up, the EEPROM on
the wiring board is defective.
Contact INFICON customer service.
W62 0 EEPROM parameters
lost!
Changed parameters were detected
during the running up.
The affected parameters are listed in
the warning.
Check the settings of the listed
parameters.
If the message occurs constantly
during running up, the EEPROM on
the wiring board is defective.
Contact INFICON customer service.