INFICON Maintenance | 8
Ecotec E3000-Operating-instructions-kina22en1-28-(2211) 73 / 104
Fig.42: Entries in a maintenance list
The detailed view also shows maintenance work that you as operator cannot confirm.
In order to explain and confirm this work, you have to have special technical
knowledge and access to the “Service” menu.
Confirm maintenance In the “Confirm maintenance” menu, confirm the replacement of the oil wick cartridge
and the air filter replacement.
Select “Oil wick cartridge” or “Air filter” and then press “OK.” You are then prompted to
state whether you want to confirm the maintenance, i.e. make an entry in the
maintenance list.
The maintenance intervals for the oil wick cartridge and the air filter of the basic unit
are specified and the system will remind you when the interval has elapsed.
Sniffer filter
Because the device can be operated with different sniffer lines, any replacement of the
filter in the sniffer tip is not subject to the maintenance schedule. Instead, you can
enter a period in “Sniffer filter” after which the device will ask you to replace the filter.
Setting range: 10 to 1000hours and infinite (∞).
Set ∞ if you do want to suppress the request for replacement.
Fig.43: Confirming maintenance work
Maintenance plan If you select “Maintenance schedule,” you have the option of deactivating the plan and
thus the reminder messages. Under the maintenance schedule menu item, the
warning W80 "Cathode switched over!" can be deactivated irrespective of the other
reminders.