INFICON Installation | 5
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 39 / 152
WARNING
Fire hazard due to overload, short circuit and overheating
► Observe the permissible ambient conditions.
► Ensure sufficient ventilation, especially on the ventilation slots on the left and right
of the device: There should be free space in the front, to the rear and sides of the
unit of at least 10 cm.
► Do not block the fan inlet in the accessory compartment.
► Keep heat sources away from the device.
► Make sure that the power switch or power plug is always easily accessible.
► Disconnect the device from the mains immediately if smoke is emitted from the
device.
WARNING
Device may fall over due to earthquake
► In areas where there is a risk of earthquakes, secure the device to the device
handle with a short chain or a tension belt (designed for 1000 N).
Fig.7:
Handle of the device
CAUTION
Risk of injury due to uncontrolled rolling away
► To ensure proper functioning of the parking brakes, remove the adhesive strips
from the wheels.
► To avoid uncontrolled rolling away, fix the device by locking the wheels with parking
brake.