6 | Operation INFICON
76 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
opacity (grayed out) to normal opacity. To make the background display
visible in standby mode, activate the corresponding function, see “Change
general display settings [}58]”.
ð Inlet area
Additionally to the deduction of the internal background, pressing START with
every measurement also deducts the background of the inlet area.
IF you would like to switch this on, make sure that the background of the inlet
area is determined, see "Determine the background of the inlet area [}76]".
We recommend that the calculation of the inlet area background is
occasionally repeated, because the device pumps down any background
present in the inlet area after a certain time.
ð Switched off
No background suppression.
3
Save .
High background value
If the actual background is higher than 1×10
-8
mbar l/s, it can no longer be
automatically deducted. The high background must be pumped down in this case.
6.3.13.1 Determine the background of the inlet area
This function determines the value of the internal Helium background.
If "Zero and Filter" are shown in the window under background suppression with value
"Inlet area" selected, this value is then deducted from the measured signal
whenSTART is pressed. See also "Change background suppression [}75]".
ü Inlet blank flange
ü Operation mode Vacuum Mode
ü The device is in the state "Standby" or "Measure".
1
Select the cross-hair icon .
2
Select the calibration "Inlet background”.
3
To start the automatic determination of the intake background, press .
6.3.14 Change SL3000 settings for sound / light
You can set up how you are to be additionally informed on the display of the
touchscreen concerning the measurements. The settings take effect with Setpoint 1.
ü
Operator
or
Supervisor
rights
1
> Setup > Measurement > Sniffing > SL3000 sniffer line
2
Select "Sniffer line light alarm configuration".
ð As choices there are "Lighter", Off" or "Flash".