2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-3
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in Table
6-1.
Table 6-1
Compliance limits
2400 2410 2420
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
200mV
2V
20V
200V
±210mV
±2.1V
±21V
±210V
200mV
2V
20V
1000V
±210mV
±2.1V
±21V
±1.1kV
200mV
2V
20V
60V
±210mV
±2.1V
±21V
±63V
1µA
10µA
100µA
1mA
10mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
1µA
10µA
100µA
1mA
20mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±21mA
±105mA
±1.05A
10µA
100µA
1mA
10mA
100mA
1A
3A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±3.15A
2425/2430 2440
Measure
range
Maximum
compliance
value
Measure
range
Maximum
compliance
value
200mV
2V
20V
100V
±210mV
±2.1V
±21V
±105V
200mV
2V
10V
40V
±210mV
±2.1V
±10.5V
±42V
10µA
100µA
1mA
10mA
100mA
1A
3A/10A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
*
10µA
100µA
1mA
10mA
100mA
1A
5A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±5.25A
* ±3.15A (2425 and 2430 DC mode)
±10.5A (Pulse mode; 2430 only)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals