EasyManuals Logo

Keithley SourceMeter 2420 User Manual

Keithley SourceMeter 2420
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #325 background imageLoading...
Page #325 background image
13-12 Digital I/O Port, Safety Interlock, and Output Configuration 2400 Series SourceMeter
®
User’s Manual
Output configuration programming example
Table 13-3 lists the command sequence for output configuration. These commands set up
the SourceMeter as follows:
Interlock: enabled
Output-off mode: hi impedance
Auto-off mode: on
NOTE Connect pins 8 and 9 of the digital I/O port together to simulate a closed inter-
lock switch. Otherwise, the unit will not turn on its output when the measure-
ment is made.
Table 13-3
Output configuration programming example
Command Description
*RST
:SOUR:VOLT 10
:OUTP:INT:STAT ON
:OUTP:SMOD HIMP
:SOUR:CLE:AUTO ON
:READ?
Restore GPIB defaults.
Output 10V.
Enable interlock.*
Select high impedance output-off mode.
Enable auto-off mode.
Trigger and acquire readings.
*Connect pins 8 and 9 of digital I/O port to simulate closed interlock switch.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals