2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-21
Table 18-10
TRACe command summary
Command Description
Default
parameter SCPI
:TRACe|:DATA
:DATA?
:CLEar
:FREE?
:POINts <NRf>
:ACTual?
:POINts?
:FEED <name>
:CONTrol <name>
:CONTrol?
:TSTamp
:FORMat <name>
:FORMat?
Use :TRACe or :DATA as root command:
Read the contents of the buffer (data store).
Clear readings from buffer.
Query bytes available and bytes in use.
Specify size of buffer (1 to 2500).
Queries number of readings stored in the buffer.
Query buffer size.
Select source of readings (SENSe[1],
CALCulate[1], or CALCulate2).
Specify buffer control mode (NEVER or NEXT).
Query buffer control mode.
Path to set timestamp format:
Select format (ABSolute or DELTa).
Query timestamp format.
(Note)
Note: :SYSTem:PRESet and *RST have no effect on the commands in this subsystem.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals