F-18 Contact Check Function 2400 Series SourceMeter
®
User’s Manual
FAIL?
:CALCulate2:LIMit4:FAIL? Read LIMIT 4 test result
Description This command is used to read the results of the LIMIT 4 test:
0 = Limit 4 test passed
1 = Limit 4 test failed
Reading the results of the limit test does not clear the fail indication of
the test. A failure can be cleared by sending a :SYSTem:CLEar
command.
SOURce2 <NRf> |<NDN>
:CALCulate2:LIMit4:SOURce2 <NRf> |<NDN> Specify pattern; LIMIT 4 failure
Parameters <NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value
Query :SOURce2? Query source value for specified limit
Description This command is used to define the digital output fail patterns for the
Limit 4 test (0 to 7, 3-bit; 0 to 15, 4-bit).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals