2400 Series SourceMeter
®
User’s Manual Digital I/O Port, Safety Interlock, and Output Configuration 13-5
Remote digital output control
Use the :SOURce2:TTL <NRf> command to control the digital output line logic levels,
where <NRf> is the decimal value shown in Table 13-1. For example, send the following
command to set the output lines to L, H, L, H:
:SOUR2:TTL 5
Safety interlock
The digital I/O port provides an interlock line for use with a test fixture interlock switch.
When properly used, the OUTPUT of the SourceMeter will turn OFF when the lid of the
test fixture is opened. See Section 2, “Connections,” for important safety information
when using the test fixture interlock.
Table 13-1
Digital output line settings
OUT 4 OUT 3 OUT 2 OUT 1
Decimal
value*
L
L
L
L
L
L
L
L
H
H
H
H
H
H
H
H
L
L
L
L
H
H
H
H
L
L
L
L
H
H
H
H
L
L
H
H
L
L
H
H
L
L
H
H
L
L
H
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
L = Low (Gnd)
H = High (>+3V)
*0-7 in 3-bit mode, which is controlled by CONFIG LIMIT menu.
(See Section 12.) 0-65535 with 2499-DIGIO 16-bit option.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals