2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-79
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or MAXi-
mum parameters on a lower source range will generate error -221 (Set-
ting Conflict).
TRIGgered[:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude] <n> Set fixed I-Source amplitude when
triggered
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n> Set fixed V-Source amplitude when
triggered
Parameters 2400
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
-210 to 210 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -1.05A or -210V
MAXimum +1.05A or +210V
2410
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
-1100 to 1100 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -1.05A or -1100V
MAXimum +1.05A or +1100V
2420
<n> = -3.15 to 3.15 Set I-Source amplitude (amps)
-63 to 63 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -3.15A or -63V
MAXimum +3.15A or +63V
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Set I-Source amplitude (amps)
-105 to 105 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -3.15A or -105V
MAXimum +3.15A or +105V
2430 Pulse Mode
<n> = -10.5 to 10.5 Set I-Source amplitude (amps)
-105 to 105 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -10.5A or -105V
MAXimum +10.5A or +105V
2440
<n> = -5.25 to 5.25 Set I-Source amplitude (amps)
-42 to 42 Set V-Source amplitude (volts)
DEFault 0A or 0V
MINimum -5.25A or -42V
MAXimum +5.25A or +42V