EasyManuals Logo

Keithley SourceMeter 2420 User Manual

Keithley SourceMeter 2420
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #170 background imageLoading...
Page #170 background image
2400 Series SourceMeter
®
User’s Manual Source-Measure Concepts 6-27
Figure 6-12
V-Source output characteristics
210V21V
105mA
1.05A
A. Model 2400
1100V21V
21mA
1.05A
B. Model 2410
63V21V
1.05A
3.15A
C. Model 2420
21V
1.05A
3.15A
D. Models 2425 and 2430
10.5A
Model 2430
Pulse Mode
DC Mode
105V
Limit I Limit I
Limit I
Limit I
Source V
Source V
Source VSource V
42V10.5V
1.05A
5.25A
E. Model 2440
Limit I
Source V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals