10-20 Sweep Operation 2400 Series SourceMeter
®
User’s Manual
4. Use the displayed menu selections to enter the desired # POINTS, individual point
values (ADJUST POINTS), and INIT (initial) value.
5. From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press ENTER,
then choose FINITE or INFINITE as desired.
6. Again from the CONFIGURE SWEEPS menu, choose SOURCE RANGING,
press ENTER, then select BEST FIXED, AUTO RANGE, or FIXED as
appropriate.
7. Press EXIT to return to normal display.
Step 3: Set delay.
Set the source delay as follows:
1. Press CONFIG then SOURCE V or SOURCE I depending on the selected source
function.
2. Select DELAY, then press ENTER.
3. Set the delay to the desired value, then press ENTER.
4. Press EXIT to return to normal display.
Step 4: Turn output on.
Press the ON/OFF OUTPUT key to turn the output on (OUTPUT indicator turns on). The
SourceMeter will output the programmed bias level.
Step 5: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off
by pressing the ON/OFF OUTPUT key.
Step 6: Read buffer.
Use the RECALL key to access the source-measure readings stored in the buffer. Use the
TOGGLE to display statistical information.