Page #434 background imageLoading...
Page #434 background image
2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-53
Sorting mode status bit values:
Result
1
Bit #: 21 20 19 9 8
Meas. Event
Status
2
Limit 1 and 4 pass and 2, 3, and
5-12 disabled
0 0 0 0 0 Bit 5 (LP)
Limit test 1 fail 0 0 0 0 1 Bit 0 (L1)
Limit test 2 pass 0 0 0 1 0 Bit 5 (LP)
Limit test 3 pass 0 0 0 1 1 Bit 4 (HL3)
Limit test 4 fail 0 0 1 0 1 Bit 10 (CC)
Limit test 5 pass 0 0 1 0 0 Bit 5 (LP)
Limit test 6 pass 0 0 1 1 0 Bit 5 (LP)
Limit test 7 pass 0 0 1 1 1 Bit 5 (LP)
Limit test 8 pass 0 1 0 0 0 Bit 5 (LP)
Limit test 9 pass 0 1 0 0 0 Bit 5 (LP)
Limit test 10 pass 0 1 0 1 0 Bit 5 (LP)
Limit test 11 pass 0 1 0 1 1 Bit 5 (LP)
Limit test 12 pass 0 1 1 0 0 Bit 5 (LP)
Limit 1 and 4 pass and 2, 3, and
5-12 fail
1 1 1 1 1 -
1. Limit 4 test available only with contact check option (Appendix F).
2. See Section 15, Measurement Event Register” and Figure 15-6 for details.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals