18-76 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
2425 and 2430 DC Mode
<n> = -3.15 to 3.15 Specify I-Source level (amps)
-105 to 105 Specify V-Source level (volts)
DEFault 100µA range (I-Source)
20V range (V-Source)
MINimum 10µA range (I-Source)
200mV range (V-Source)
MAXimum 3A range (I-Source)
100V range (V-Source)
UP Select next higher range
DOWN Select next lower range
2430 Pulse Mode
<n> = -10.5 to 10.5 Specify I-Source level (amps)
-105 to 105 Specify V-Source level (volts)
DEFault 100µA range (I-Source)
20V range (V-Source)
MINimum 10µA range (I-Source)
200mV range (V-Source)
MAXimum 10A range (I-Source)
100V range (V-Source)
UP Select next higher range
DOWN Select next lower range
2440
<n> = -5.25 to 5.25 Specify I-Source level (amps)
-42 to 42 Specify V-Source level (volts)
DEFault 100µA range (I-Source)
10V range (V-Source)
MINimum 10µA range (I-Source)
200mV range (V-Source)
MAXimum 5A range (I-Source)
40V range (V-Source)
UP Select next higher range
DOWN Select next lower range
Query :RANGe? Query range for specified source
:RANGe? DEFault Query *RST default source range
:RANGe? MINimum Query lowest source range
:RANGe? MAXimum Query highest source range
Description This command is used to manually select the range for the specified
source. Range is selected by specifying the approximate source magni-
tude that you will be using. The instrument will then go to the lowest
range that can accommodate that level. For example, if you expect to
source levels around 3V, send the following command:
:SOURce:VOLTage:RANGe 3