3. Basic operation
MKC-710 25
To change the life value of anolyte to zero, press [Reagent][Anode][Life]
current life value[0]to[ENTER]key
When use the Auto Solvent Change Unit
injection
1) Press [MENU/HOME] key.
Select “1.Manual Operation” and press
[ENTER] key. Select [Pump].
Press Injection Pump [On] key to
inject approx. 100mL of anolyte into
the titration cell. (The lower line of the
titration cell is a 100mL mark.)
[OffOn] key automatically changes
to [OnOff] at this time.
Press Injection Pump again, and
injection will stop.
2) After changing anolyte, set the life value of anolyte to 0.